Search results for "Pulse processing"
showing 10 items of 21 documents
Digital performance improvements of a CdTe pixel detector for high flux energy-resolved X-ray imaging
2015
Abstract Photon counting detectors with energy resolving capabilities are desired for high flux X-ray imaging. In this work, we present the performance of a pixelated Schottky Al/p-CdTe/Pt detector (4×4) coupled to a custom-designed digital readout electronics for high flux measurements. The detector (4×4×2 mm 3 ) has an anode layout based on an array of 16 pixels with a geometric pitch of 1 mm (pixel size of 0.6 mm). The 4-channel readout electronics is able to continuously digitize and process the signals from each pixel, performing multi-parameter analysis (event arrival time, pulse shape, pulse height, pulse time width, etc.) even at high fluxes and at different throughput and energy re…
A digital approach for real time high-rate high-resolution radiation measurements
2014
Abstract Modern spectrometers are currently developed by using digital pulse processing (DPP) systems, showing several advantages over traditional analog electronics. The aim of this work is to present digital strategies, in a time domain, for the development of real time high-rate high-resolution spectrometers. We propose a digital method, based on the single delay line (SDL) shaping technique, able to perform multi-parameter analysis with high performance even at high photon counting rates. A robust pulse shape and height analysis (PSHA), applied on single isolated time windows of the detector output waveforms, is presented. The potentialities of the proposed strategy are highlighted thro…
Performance of a digital CdTe X-ray spectrometer in low and high counting rate environment
2010
Abstract The high performances of CdTe detectors for X-ray and gamma ray spectroscopy are already well known. Among the traditional semiconductor spectrometers, CdTe detectors show high detection efficiency and good room temperature performance and are well suited for the development of compact detection systems. In this work, we investigated the performance of a CdTe detector coupled with a custom digital pulse processing (DPP) system for X-ray spectroscopy. The DPP method, implemented on a PC platform, performs a pile-up inspection and a pulse height analysis of the preamplifier output pulses, digitized by a 14-bit, 100 MHz ADC. The spectroscopic results point out the excellent performanc…
High-rate dead-time corrections in a general purpose digital pulse processing system
2015
The abilities on dead-time correction of a real-time digital pulse processing (DPP) system for high-rate high-resolution radiation measurements are presented. The DPP system, through a fast and slow analysis of the output waveform from radiation detectors, is able to perform an accurate estimation of the true input counting rate (ICR), a fine pulse height (energy) and shape (peaking time) analysis even at high ICRs.
Direct Measurement of Mammographic X-Ray Spectra with a Digital CdTe Detection System
2012
In this work we present a detection system, based on a CdTe detector and an innovative digital pulse processing (DPP) system, for high-rate X-ray spectroscopy in mammography (1–30 keV). The DPP system performs a height and shape analysis of the detector pulses, sampled and digitized by a 14-bit, 100 MHz ADC. We show the results of the characterization of the detection system both at low and high photon counting rates by using monoenergetic X-ray sources and a nonclinical X-ray tube. The detection system exhibits excellent performance up to 830 kcps with an energy resolution of 4.5% FWHM at 22.1 keV. Direct measurements of clinical molybdenum X-ray spectra were carried out by using a pinhole…
Digital Techniques for High-Rate High-Resolution Radiation Measurements
2014
Digital pulse processing (DPP) techniques are increasingly used in the development of modern spectroscopic systems. DPP systems, based on direct digitizing and processing of detector signals (preamplifier output signals), ensure higher flexibility, stability, lower dead time, higher throughput and better energy resolution than traditional pulse processing systems. In this work, we present our progress in the development of DPP systems for high-rate high-resolution radiation measurements. An innovative digital system, able to perform multi-parameter analysis (input counting rate, pulse height, pulse shape, event arrival time, etc.) even at high photon counting rates is presented. Experimenta…
Experimental results from Al/p-CdTe/Pt X-ray detectors
2013
Recently, Al/CdTe/Pt detectors have been proposed for the development of high resolution X-ray spectrometers. Due to the low leakage currents, these detectors allow high electric fields and the pixellization of anodes with the possibility to realize single charge carrier sensing detectors. In this work, we report on the results of electrical and spectroscopic investigations on CdTe diode detectors with Al/CdTe/Pt electrode configuration (4.1×4.1×0.75 and 4.1×4.1×2 mm3). The detectors are characterized by very low leakage currents in the reverse bias operation: 0.3 nA at 25 °C and 2.4 pA at -25 °C under a bias voltage of -1000 V. The spectroscopic performance of the detectors at both low and…
Vers une architecture électronique unifiée et zéro temps mort pour l'instrumentation nucléaire
2014
International audience; Dans l'instrumentation nucléaire, et d'autres domaines tels que les applications de la physique des particules, l'analyse de signaux médicaux (Tomographie par Émission de Positons (TEP), Électro-EncéphaloGraphie (EEG)) ou la détection de signaux radar, le signal mesuré est composé d'impulsions dont la date d'arrivée est aléatoire, et dont l'amplitude et la durée sont non déterministes. La grande variété d'applications nécessite de disposer d'une architecture flexible pouvant être aisément reprogrammée. De plus, l'exigence de mesures en temps réel impose de disposer d'une grande capacité de calcul et de bande passante pour la mise en forme et l'extraction des caractér…
High-rate x-ray spectroscopy in mammography with a CdTe detector: A digital pulse processing approach
2010
Purpose:Direct measurement of mammographic x-ray spectra under clinical conditions is a difficult task due to the high fluence rate of the x-ray beams as well as the limits in the development of high resolution detection systems in a high counting rate environment. In this work we present a detection system, based on a CdTe detector and an innovative digital pulse processing (DPP) system, for high-rate x-ray spectroscopy in mammography. Methods: The DPP system performs a digital pile-up inspection and a digital pulse height analysis of the detector signals, digitized through a 14-bit, 100 MHz digitizer, for x-ray spectroscopy even at high photon counting rates. We investigated on the respon…
Instrumentation for time-of-flight elastic recoil detection analysis
2016
Time-of-flight elastic recoil detection is an ion beam based method to analyze the elemental composition of thin film samples at different depths. In order to improve the mass resolution and to enable kinematic correction a position sensitive gas ionization chamber energy detector was constructed. This detector along with pre-existing timing detectors were connected to a modern fully digitizing data acquisition setup. This thesis describes the design of these instruments, including all aspects from the mechanical design of the gas ionization detector to the algorithm and software development of the digitizing acquisition setup. The performance of the system has been studied with measurements …